What Should Your Dental Office Receive After an X-Ray EPE?

The evaluation is complete. Here’s what should happen next.

​

When your dental X-ray equipment has its required Equipment Performance Evaluation (EPE)—often called an X-ray calibration—the testing is only part of the process. Your office also needs the documentation showing what was evaluated, the results, and any corrective action that may be necessary. Here’s what to expect.

​

1. A Complete EPE Report

Your office should receive documentation identifying the X-ray equipment evaluated and the results of the required performance testing.

Keep the completed EPE with your office’s radiation safety records. At North Star Dental Services, EPE reports are sent electronically through DocuSign. Your office signs to acknowledge receipt of the completed EPE, and the signed document is automatically returned to North Star. The signature does not mean you're expected to independently verify the technical measurements. That's the responsibility of the qualified person performing the evaluation.

 

2. An Updated Technique Chart

Your office should also have a current technique chart appropriate for the X-ray equipment and imaging procedures being performed.

The technique chart should be displayed on the X-ray unit or near its controller where operators can readily reference it.

North Star provides equipment-specific technique charts for:

  • Intraoral X-ray systems

  • Panoramic systems

  • Cephalometric systems

  • CBCT systems

 

Technique charts can disappear during remodeling, equipment moves, cleaning, or simply over time. After an EPE is a good time to make sure yours is still where it belongs.

 

3. Clear Communication About Anything That Didn't Pass

If a required test doesn't pass, that shouldn't simply appear as a note buried somewhere in the report. Your office should understand what failed and what corrective action needs to happen. A failed EPE requires prompt attention and may require repair or replacement of the equipment.

Ask:

  • Which test did not pass?

  • What does that mean for continued use of the equipment?

  • Can the problem be repaired, or does the unit need to be replaced?

  • Is additional testing required after the corrective action?

  • What documentation should we retain?

 

Related: What Happens When Your Dental X-Ray Doesn't Pass Its EPE?

 

4. Additional Documentation for New Equipment

If the X-ray equipment was newly installed, additional records may apply. When North Star installs new X-ray equipment, we provide applicable FDA Form 2579 documentation along with manufacturer warranty information. Keep these documents with the equipment's permanent records.

​

Before You File It Away

You don't need to become an X-ray testing expert. But before putting the paperwork in your Radiation Safety Manual or electronic records, make sure:

The report belongs to your office and the correct X-ray unit.
You received the complete EPE documentation.
The appropriate person signed to acknowledge receipt.
The current technique chart is displayed.
Any failed test has been discussed and corrective action is underway.

Then keep the completed documentation where your Radiation Safety Officer can find it.

​